GC
- Samples
- Benzene , Toluene , Ethylbenzene , p-Xylene , m-Xylene , iso-Propylbenzene , o-Xylene , n-Propylbenzene , p-Ethyltoluene , m-Ethyltoluene , o-Ethyltoluene , 1,2,5-Trimethylbenzene , 1,2,4-Trimethylbenzene , 1,2,3-Trimethylbenzene , Naphthalene
- Applications
- Aromatic hydrocarbons in Regular gasoline
- Column
- ULBON HR-20M
- Column Size
- 50m×0.25mm(0.5μm)
- Column temp.
- 50°C→180°C(2°C/min)
- Carrier gas
- He,1.04ml/min
- Detector temp.
- 250°C(FID)
- Injection volume
- 0.2μl