GC
- Samples
- Benzene , Toluene , Ethylbenzene , Isopropylbenzene , n-Propylbenzene , Styrene monomer , Cyclopentanol , N, N-Dimethylformamide ,
- Applications
- Measurement of styrene monomer in DMF using Shinwasorb
- Column
- FFAP 10% Shinwasorb-U 60/80
- Column Size
- 3.1m×3.2mm glass
- Column temp.
- 120°C
- Carrier gas
- 50 mL/min N2
- Injection port temp.
- 240°C
- Detector temp.
- 240°C
- Detector
- FID
- Injection volume
- 1.0 μL