GC
- Samples
- Diazinon, Di-n-butyl phthalate in Acetone
- Applications
- Analysis of diazinon using Shinwasorb
- Column
- Silicone OV-11 10% Shinwasorb-U 60/80
- Column Size
- 1.6 m x 3.2 mm I.D. Glass
- Column temp.
- 200℃
- Carrier gas
- 25 mL/min N₂(Adjust so that the retention time of diazinon is about 7 minutes.)
- Injection port temp.
- 230°C
- Detector temp.
- 230°C
- Detector
- FID
- Injection volume
- 3.0 μL