LC
- Samples
- Predonisone , Cortisone , Hydrocortisone , Cortisone acetate , Hydrocortisone acetate
- Applications
- Separation of the steroid
- Column
- ULTRON VX-ODS
- Column Size
- 150 mm L × 4.6 mm I.D.
- Detector
- UV-254 nm
- Particle Size
- 5 µm
- Mobile phase
- H2O/CH3OH/THF = 70/10/20
- Frow rate
- 1.0 mL/min
- Column temp.
- 40°C