Search result
"Sunpak-A" Filtered 21 Item.
GC
- Samples
- Ethanol , Isopropyl Alcohol , Methyl Ethyl Ketone
- Applications
- Lower alcohol in water and Ketones
- Column
- Sunpak-A
- Column Size
- 2.1m×3.2mm
- Column temp.
- 180°C
- Carrier gas
- N2,60ml/min
- Injection port temp.
- 240°C
- Detector temp.
- 240°C(FID)
View detail
GC
- Samples
- Methane , Carbon Dioxide , Ethylene , Ethane , Water
- Applications
- Methane CO2 Ethylene Ethane
- Column
- Sunpak-A
- Column Size
- 2.1m×3.2mm
- Column temp.
- 50°C
- Carrier gas
- He,50ml/min
- Injection port temp.
- 150°C
- Detector temp.
- 150°C(TCD)
View detail
GC
- Samples
- Ethanol , Acetonitrile , Isopropanol , n-Hexane
- Applications
- Lower alcohol in water and Hydrocarbon
- Column
- Thermon-1000 5% Sunpak-A
- Column Size
- 2.1m×3.2mm
- Column temp.
- 180°C
- Carrier gas
- N2,50ml/min
- Injection port temp.
- 250°C
- Detector temp.
- 230°C(FID)
View detail
GC
- Samples
- Methanol , Acetaldehyde , Ethanol , Isopropanol , n-Propanol , Ethyl Acetate , Isobutanol , n-Butanol , Isoamyl Alcohol
- Applications
- Lower alcohol in water
- Column
- Thermon-1000 5% Sunpak-A
- Column Size
- 2.1m×3.2mm
- Column temp.
- 180°C
- Carrier gas
- N2,50ml/min
- Injection port temp.
- 240°C
- Detector temp.
- 240°C(FID)
View detail
GC
- Samples
- IPA , Methyl Cellosolve , n-Butanol , 1-Methoxy-2-Propanol , Ethyl Cellosolve , Butyl Cellosolve
- Applications
- Lower alcohol and Cellosolves
- Column
- Thermon-1000 5% Sunpak-A
- Column Size
- 2.1m×3.2mm
- Column temp.
- 170°C→220°C(6°C/min)
- Carrier gas
- N2,50ml/min
- Injection port temp.
- 260°C
- Detector temp.
- 260°C(FID)
View detail
GC
- Samples
- Acrylonitrile , Methyl Methacrylate Monomer , DMF , Chlorobenzene , Styrene Monomer , α-Methylstyrene Monomer , N,N-Dimetylformamide
- Applications
- Residual solvent in the packaging material
- Column
- Thermon-1000 5% Sunpak-A
- Column Size
- 2.1m×3.2mm
- Column temp.
- 220°C
- Carrier gas
- N2,50ml/min
- Injection port temp.
- 260°C
- Detector temp.
- 260°C
- Detector
- FID
- Injection volume
- 1 μL
View detail
GC
- Samples
- Methyl Alcohol , Acetaldehyde , Ethyl Alcohol , n-Propyl Alcohol , Isobutyl Alcohol , Isoamyl Alcohol
- Applications
- Aldehyde in the wine and Lower alcohol
- Column
- Thermon-1000 5% Sunpak-A
- Column Size
- 2.1m×3.2mm
- Column temp.
- 140°C→180°C(6°C/min)
- Carrier gas
- N2,60ml/min
- Injection port temp.
- 260°C
- Detector temp.
- 260°C(FID)
View detail
GC
- Samples
- Monomethylamine , Dimethylamine , Trimethylamine
- Applications
- Tri-, di-, monomethylamine in Nitrogen
- Column
- Thermon-3000+KOH(5%+1%) Sunpak-A
- Column Size
- 2.1m×3.2mm
- Column temp.
- 150°C
- Carrier gas
- N2,50ml/min
- Injection port temp.
- 240°C
- Detector temp.
- 240°C(FID)
View detail
GC
- Samples
- Methylamine , Dimethylamine , Trimethylamine
- Applications
- Tri-, di-, monomethylamine in water
- Column
- Thermon-3000+KOH(5%+1%) Sunpak-A
- Column Size
- 2.1m×3.2mm
- Column temp.
- 140°C
- Carrier gas
- N2,50ml/min
- Injection port temp.
- 250°C
- Detector temp.
- 230°C(FID)
View detail
GC
- Samples
- Monomethylamine , Dimethylamine , Trimethylamine
- Applications
- Tri-, di-, monomethylamine in water
- Column
- Thermon-3000+KOH(5%+1%) Sunpak-A
- Column Size
- 2.1m×3.2mm
- Column temp.
- 140°C
- Carrier gas
- N2,50ml/min
- Injection port temp.
- 260°C
- Detector temp.
- 260°C(FID)
View detail
GC
- Samples
- Propyleneglycol , 1,3-propanediol
- Applications
- Measurement of food additives using Sunpak-A
- Column
- Sunpak-A 50/80
- Column Size
- 2.1 m x 3.2 mm I.D. Glass
- Column temp.
- 180℃
- Carrier gas
- 50 mL/min N₂
- Injection port temp.
- 200℃
- Detector temp.
- 230℃
- Detector
- FID
- Injection volume
- 1.0 μL
View detail
GC
- Samples
- Propyleneglycol , 1,3-propanediol
- Applications
- Measurement of food additives using Sunpak-A
- Column
- Sunpak-A 50/80
- Column Size
- 2.1 m x 3.2 mm I.D. Glass
- Column temp.
- 180℃
- Carrier gas
- 50 mL/min N₂
- Injection port temp.
- 200℃
- Detector temp.
- 230℃
- Detector
- FID
- Injection volume
- 1.0 μL
View detail
GC
- Samples
- water, acetonitrile in methanol
- Applications
- Analysis of panipenem using Sunpak-A ~Water~
- Column
- Sunpak-A 80/100
- Column Size
- 2.1m x 3.2 mm I.D. Glass
- Column temp.
- 125℃
- Carrier gas
- 29 mL/min He (Adjust so that the retention time of acetonitrile is about 8 minutes.)
- Injection port temp.
- 200°C
- Detector temp.
- 200°C
- Detector
- FID
- Injection volume
- 1.0 μL
View detail
GC
- Samples
- Organic solvents
- Applications
- Analysis of organic solvents using Sunpak-A -1-
- Column
- Sunpak-A 50/80
- Column Size
- 2.0 m x 3.0 mm I.D. Glass
- Column temp.
- 150℃ ~ 220℃ (3 min hold) program rate 4℃/min
- Carrier gas
- 50 mL/min N₂
- Injection port temp.
- 250°C
- Detector temp.
- 250°C
- Detector
- FID
- Injection volume
- 1.0 μL
View detail
GC
- Samples
- Organic solvents
- Applications
- Analysis of organic solvents using Sunpak-A -2-
- Column
- Sunpak-A 50/80
- Column Size
- 2.0 m x 3.0 mm I.D. Glass
- Column temp.
- 150℃ ~ 220℃ (3 min hold) program rate 4℃/min
- Carrier gas
- 50 mL/min N₂
- Injection port temp.
- 250°C
- Detector temp.
- 250°C
- Detector
- FID
- Injection volume
- 1.0 μL
View detail