"FAME" Filtered 6 Item.

Graph

GC

Samples
long-chain fatty acid methyl esters (FAMEs)
Applications
Measurement of long-chain fatty acid methyl esters (FAMEs C10~C24) using Shinwasorb
Column
Shinchrom E71 5% Shinwasorb-U 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
230℃
Carrier gas
40 mL/min N₂
Injection port temp.
250℃
Detector temp.
250℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
palm oil
Applications
Measurement of fatty acid methyl esters (palm oil) using Shinwasorb
Column
Shinchrom E71 5% Shinwasorb-U 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
210℃
Carrier gas
40 mL/min N₂
Injection port temp.
260℃
Detector temp.
260℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
beef tallow
Applications
Measurement of Measurement of beef tallow fatty acid methyl esters (FAMEs) using Shinwasorb
Column
Shinchrom E71 5% Shinwasorb-U 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
210℃
Carrier gas
40 mL/min N₂
Injection port temp.
260℃
Detector temp.
260℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Palmitic acid , Stearic acid , Oleic acid , Linoleic acid , Linolenic acid
Applications
Measurement of fatty acid methyl esters (soy bean oil) using Shinwasorb
Column
Shinchrom E71 5% Shinwasorb-U 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
230℃
Carrier gas
40 mL/min N₂
Injection port temp.
260℃
Detector temp.
260℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

LC

Samples
Acesulfame K , Saccharin , Caffeine , Sorbic acid , Benzoic acid , Quinine , Aspartame
Applications
Analysis of food additives
Column
ULTRON AF-HILIC-CD
Column Size
150 mm L × 3.0 mm I.D.
Mobile phase
A: 20mM CH3CO2NH4 (pH4.7) /CH3CN = 10/90, B: 20mM CH3CO2NH4 (pH4.7) /CH3CN = 70/30
Frow rate
Refer to the picture
Column temp.
30°C
Injection volume
2.0 µL
Diluent
H2O /CH3CN = 50/50

View detail


Graph

LC

Samples
Acesulfame K , Saccharin , Sucralose , Advantame , Neotame , Aspartame
Applications
Analysis of artificial sweetener
Column
ULTRON AF-HILIC-CD
Column Size
150 mm L × 4.6 mm I.D.
Mobile phase
100mM CH3CO2NH4 (pH6.8) /CH3CN = 10/90
Frow rate
1.4 mL/min
Column temp.
30°C
Injection volume
4.0 µL
Diluent
H2O /CH3CN = 50/50

View detail