Shinwa Chemical Industries Ltd.
LC/GC Search / Application Note

Shinwa Chemical Industries
Application Search

Search result
"Fenitrothion" Filtered 5 Item.

Graph

GC

Samples
Chloroform , Fenitrothion , Fluoranthene
Applications
Analysis of fenitrothion using Shinwasorb
Column
Polyphenyl Ether (5 rings) 1.5% Shinwasorb-S 100/120
Column Size
1.1 m x 3.2 mm I.D. Glass
Column temp.
195℃
Carrier gas
55 mL/min N₂
Injection port temp.
230℃
Detector temp.
230℃
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Chloroform , Fenitrothion , Fluoranthene
Applications
Analysis of fenitrothion using Shinwasorb
Column
Polyphenyl Ether (5 rings) 1.5% Shinwasorb-S 100/120
Column Size
1.1 m x 3.2 mm I.D. Glass
Column temp.
195℃
Carrier gas
55 mL/min N₂ (Adjust so that the retention time of fenitrothion is about 16 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Anhydrous caffeine, Fenitrothion in Acetone
Applications
Analysis of Fenitrothion Emulsifiable Concentrate using Shinwasorb
Column
Silicone OV-210 3% Shinwasorb-S 80/100
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
175℃
Carrier gas
51 mL/min N₂(Adjust so that the retention time of Anhydrous caffeine is about 10 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Anhydrous caffeine, Fenitrothion in Acetone
Applications
Analysis of Fenitrothion Microcapsule using Shinwasorb
Column
Silicone OV-210 1.5% Shinwasorb-U 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
165℃
Carrier gas
50 mL/min N₂(Adjust so that the retention time of diazinon is about 17 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Benzyl salicylate, Fenitrothion in Acetone
Applications
Analysis of fenitrothion, phthalthrin mixed emulsifiable concentrate using Shinwasorb -1-
Column
Silicone OV-210 1.5% Shinwasorb-U 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
175℃
Carrier gas
52 mL/min N₂ (Adjust so that the retention time of fenitrothion is about 15 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
2.0 μL

View detail