Search result
"Air" Filtered 11 Item.
GC
- Samples
- Carbonyl Sulfide , Hydrogen Sulfide , Carbon Disulfide , Methyl Mercaptan , Ethyl Mercaptan , Dimethyl Sulfide , Benzene(Solvent) , Dimethyl Disulfide
- Applications
- Sulfur Compounds of Air Pollution
- Column
- 1,2,3-TCEP 25% Shimalite 80-100 (AW-DMCS)
- Column Size
- 3.1m×3.2mm
- Column temp.
- 70°C
- Carrier gas
- N2,50ml/min
- Injection port temp.
- 180°C
- Detector temp.
- 180°C
- Detector
- FPD
View detail
GC
- Samples
- Air , Carbon Monoxide , Methane , Carbon Dioxide
- Applications
- Gas (Air, CO, CH4, CO2)
- Column
- Activated Charcoal 60-80
- Column Size
- 1.6m×3.2mm
- Column temp.
- 50°C
- Carrier gas
- He,50ml/min
- Injection port temp.
- 50°C
- Detector temp.
- 50°C(TCD)
View detail
GC
- Samples
- Hydrogen Sulfide , Sulfur Dioxide , Methyl Mercaptan , Ethyl Mercaptan , Dimethyl Sulfide , Carbon Disulfide , Benzene , Dimethyl Disulfide
- Applications
- Air pollution sulfur compound (1)
- Column
- PPE-5rings 10% Shimalite TPA 60-80
- Column Size
- 3.1m×3.2mm
- Column temp.
- 70°C
- Carrier gas
- N2,50ml/min
- Injection port temp.
- 180°C
- Detector temp.
- 180°C(FPD)
View detail
GC
- Samples
- n-Hexane , Methyl Alcohol , Acetone , Isopropyl Alcohol , Ethyl Acetate , Methyl Ethyl Ketone , Benzene , Trichloroethylene , Isobutyl Alcohol , Isobutyl Acetate , Toluene , n-Butyl Acetate , Isoamyl Alcohol , Ethylbenzene , m-Xylene , o-Xylene , n-Propylbenzene , Cyclohexanone
- Applications
- Organic solvent in the air (ppm level)
- Column
- SBS-100 10% Shimalite TPA 60-80
- Column Size
- 3.1m×3.2mm
- Column temp.
- 110°C
- Carrier gas
- N2,30ml/min
- Injection port temp.
- 200°C
- Detector temp.
- 200°C(FID)
View detail
GC
- Samples
- Air , Carbon Monoxide , Methane , Ethane , Carbon Dioxide , Ethylene
- Applications
- Gas (O2, N2, CO, Methan, Ethan, CO2, Ethylene)
- Column
- Silica Gel 60-80
- Column Size
- 1.6m×3.2mm
- Column temp.
- 50°C
- Carrier gas
- He,60ml/min
- Injection port temp.
- 50°C
- Detector temp.
- 50°C(TCD)
View detail
GC
- Samples
- Carbon Disulfide , Hydrogen Sulfide , Methyl Mercaptan ,Ethyl Mercaptan , DMS , Benzene . DMDS
- Applications
- Air pollution sulfur compound (3)
- Column
- β,β-ODPN 25% Chromosorb W 60-80 (AW-DMCS-ST)
- Column Size
- 3.1m×3.2mm
- Column temp.
- 70°C
- Carrier gas
- N2,50ml/min
- Detector temp.
- 180°C(FPD)
- Injection volume
- -
View detail
GC
- Samples
- Dichloromethane , trans-1,2-Dichloroethylene , Chloroform , 1,2-Dichloroethane , 1,1,1-Trichloroethane , Carbon tetrachloride , Trichloroethylene , Toluene , Tetrachloroethylene , Chlorobenzene , 1,1,2,2,-Tetrachloroethane , o-Dichlorobenzene
- Applications
- Measurement of chlorinated solvents in air using Shinwasorb
- Column
- SBS-300 10% Shinwasorb-S 60/80
- Column Size
- 3.1 m x 3.2 mm I.D. Glass
- Column temp.
- 100℃
- Carrier gas
- 30 mL/min N₂
- Injection port temp.
- 240℃
- Detector temp.
- 240℃
- Detector
- FID
- Injection volume
- 1.0 mL
View detail
GC
- Samples
- Air , HCl , Water
- Applications
- Analysis of Hydrogen chloride
- Column
- SF-96 15% Shimalite-F 20/80
- Column Size
- 3.1 m × 3.2 mm glass
- Column temp.
- 60°C
- Carrier gas
- He, 30 ml/min
- Injection port temp.
- 190°C
- Detector temp.
- 190°C
- Detector
- TCD
- Injection volume
- 1.0 ml
View detail
LC
- Samples
- Terbutaline
- Applications
- Bronchodilator agent
- Column
- ULTRON ES-PEPSIN
- Column Size
- 150 mm L × 6.0 mm I.D.
- Mobile phase
- 20 mM KH2PO4 (pH 6.0)
- Frow rate
- 1.0 mL/min
- Column temp.
- 25°C
- Injection volume
- 5 µL (ca. 0.25 mg/mL)
View detail
LC
- Samples
- Montelukast sodium
- Applications
- Analysis of Montelukast sodium
- Column
- ULTRON ES-OVM
- Column Size
- 150 mm L × 4.6 mm I.D.
- Mobile phase
- 20 mM KH2PO4 + H3PO4 (pH 3.0)/CH3OH = 60/40
- Frow rate
- 1.0 mL/min
- Column temp.
- 25°C
- Injection volume
- 2.0 µL
- Diluent
- methanol
View detail