Shinwa Chemical Industries Ltd.
LC/GC Search / Application Note

Shinwa Chemical Industries
Application Search

Search result
"Cyclohexane" Filtered 9 Item.

Graph

GC

Samples
Mixture of 13 organic solvents
Applications
Measurement of organic solvents using Shinwasorb -1-
Column
PEG1500 12.5% Shinwasorb-U 40/60
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
70℃
Carrier gas
40 mL/min N₂
Injection port temp.
200℃
Detector temp.
200℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Mixture of 22 organic solvents
Applications
Measurement of organic solvents using Shinwasorb -2-
Column
PEG1500 12.5% Shinwasorb-U 40/60
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
60~120℃ (4℃/min)
Carrier gas
40 mL/min N₂
Injection port temp.
200℃
Detector temp.
200℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Isopropyl iodide , Methylcyclohexane , o-xylene
Applications
Analysis of hydroxypropylcellulose (JP)
Column
ULBON HR-1
Column Size
0.53 mm I.D. x 30 m df = 3.0 µm
Column temp.
40℃ (3 min)~10℃/min~100℃~50℃/min~250℃ (3 min)
Carrier gas
52 cm/sec He
Injection port temp.
180°C
Detector temp.
280°C
Detector
FID
Injection volume
2.0 μL Split rate: 1:50

View detail


Graph

GC

Samples
Cyclohexane , N,N-dimethylaniline , Naphthalene ,
Applications
Analysis of ampicillin hydrate using Shinwasorb
Column
Silicone OV-17 1.5% Shinwasorb-S 60/80
Column Size
2.1 m x 2.6 mm I.D. Glass
Column temp.
105℃
Carrier gas
20 mL/min He(adjust so that the retention time of N,N-dimethylaniline is about 5 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Cyclohexane , Ethyl acetate , 2-propanol ,
Applications
Analysis of lenampicillin hydrochloride using Shinwasorb
Column
Quadrol 5% Shinwasorb-S 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
100℃
Carrier gas
70 mL/min N₂ (Adjust so that the retention time of cyclohexane is about 1 minute.)
Injection port temp.
160°C
Detector temp.
160°C
Detector
FID
Injection volume
4.0 μL

View detail


Graph

GC

Samples
Cyclohexane , Dicyclohexyl phthalate , Etofenprox
Applications
Analysis of etofenprox using Shinwasorb
Column
Silicone OV-225 2.5% Shinwasorb-U 60/80
Column Size
1.1m x 3.2 mm I.D. Glass
Column temp.
240℃
Carrier gas
37 mL/min N₂ (Adjust so that the retention time of dicyclohexyl phthalate is about 8 minutes.)
Injection port temp.
250°C
Detector temp.
250°C
Detector
FID
Injection volume
3.0 μL

View detail


Graph

GC

Samples
Etofenprox in Cyclohexane
Applications
Analysis of etofenprox using Shinwasorb -2- ~Purity (3) Related substances~
Column
Silicone OV-225 2.5% Shinwasorb-U 60/80
Column Size
1.1m x 3.2 mm I.D. Glass
Column temp.
240℃
Carrier gas
37 mL/min N₂
Injection port temp.
250°C
Detector temp.
250°C
Detector
FID
Injection volume
3.0 μL

View detail


Graph

LC

Samples
Sucrose , Lactose , Glucose , Xylose , Galactose , Fructose , Inositol
Applications
Sugar and alditol
Column
ULTRON PS-80P
Column Size
300 mm L × 8.0 mm I.D.
Mobile phase
H2O
Frow rate
1.0 mL/min
Column temp.
80°C
Pressure
2.0 MPa

View detail


Graph

LC

Samples
Erythritol , Xylitol , Fructose , Glucose , Sorbitol , Sucrose , Inositol , Maltose , Maltitol , Lactose , Lactitol
Applications
Analysis of saccharides and sweeteners
Column
ULTRON AF-HILIC-CD
Column Size
150 mm L × 2.0 mm I.D.
Mobile phase
A: 100mM CH3CO2NH4 (pH6.8) /CH3CN = 10/90, B: 100mM CH3CO2NH4 (pH6.8) /CH3CN = 70/30
Frow rate
0.32 mL/min
Column temp.
40°C
Injection volume
1.0 µL
Diluent
H2O /CH3CN = 50/50

View detail