Search result
"DEP" Filtered 10 Item.
GC
- Samples
- α-Pinene , β-Pinene , Limonene , Octyl Aldehyde , Nonyl Aldehyde , Citronellal , Decyl Aldehyde , Linalool , Linalyl Acetate , Isobornyl Acetate(Citronellol) , Menthol , Benzyl Acetate , Perillaldehyde , Geraniol , α-Ionone , Rose-P , C12-Alcohol , Anisaldehyde , Methyl Cinnamate , Ethyl Cinnamate , Eugenol , Heliotropin , DEP , Vanillin , Benzyl Benzoate , Benzyl Salicylate
- Applications
- The mixed sample of perfume component
- Column
- Thermon-3000 5% Chromosorb W 80-100 (AW-DMCS)
- Column Size
- 3.1m×3.2mm
- Column temp.
- 80°C→250°C(6°C/min)
- Carrier gas
- N2,60ml/min
- Injection port temp.
- 290°C
- Detector temp.
- 280°C(FID)
View detail
LC
- Samples
- Disopyramide phosphate
- Applications
- Disopyramide phosphate
- Column
- ULTRON ES-OVM
- Column Size
- 150 mm L × 4.6 mm I.D.
- Mobile phase
- 20 mM KH2PO4 + 20 mM K2HPO4 (pH 6.0)/C2H5OH = 90/10
- Frow rate
- 1.0 mL/min
- Column temp.
- 25°C
View detail
LC
- Samples
- Duloxetine hydrochloride
- Applications
- Duloxetine hydrochloride
- Column
- ULTRON ES-OVM
- Column Size
- 150 mm L × 4.6 mm I.D.
- Mobile phase
- 20 mM KH2PO4(pH 4.6)/CH3CN = 80/20
- Frow rate
- 1.0 mL/min
- Column temp.
- 25°C
View detail
LC
- Samples
- Duloxetine hydrochloride
- Applications
- Duloxetine hydrochloride
- Column
- ULTRON ES-PhCD
- Column Size
- 150 mm L × 6.0 mm I.D.
- Mobile phase
- 20 mM KH2PO4 + H3PO4(pH 3.0)/CH3CN = 65/35
- Frow rate
- 1.0 mL/min
- Column temp.
- 25°C
View detail
LC
- Samples
- Bupropion hydrochloride
- Applications
- Bupropion hydrochloride
- Column
- ULTRON ES-OVM
- Column Size
- 150 mm L × 4.6 mm I.D.
- Mobile phase
- 20 mM KH2PO4 + 20 mM K2HPO4(pH 6.0)/C2H5OH = 90/10
- Frow rate
- 1.0 mL/min
- Column temp.
- 25°C
View detail
LC
- Samples
- Duloxetine hydrochloride
- Applications
- Duloxetine hydrochloride
- Column
- ULTRON ES-OVM
- Column Size
- 100 mm L × 3.0 mm I.D.
- Mobile phase
- 20 mM KH2PO4(pH 4.6)/CH3CN = 80/20
- Frow rate
- 1.0 mL/min
- Column temp.
- 25°C
View detail
LC
- Samples
- Escitalopram oxalate
- Applications
- Escitalopram oxalate
- Column
- ULTRON ES-OVM , ULTRON ES-OVM-E
- Column Size
- 150 mm L × 4.6 mm I.D.
- Mobile phase
- 50 mM KH2PO4 + 0.2 N NaOH(pH 7.0)/CH3CN = 85/15
- Frow rate
- 0.6 mL/min
- Column temp.
- 30°C
View detail
LC
- Samples
- Escitalopram oxalate
- Applications
- Analysis of Escitalopram oxalate
- Column
- ULTRON ES-OVM
- Column Size
- 100 mm L × 3.0 mm I.D.
- Mobile phase
- 50 mM KH2PO4 + 0.2 N NaOH(pH 7.0)/CH3CN = 85/15
- Frow rate
- 0.6 mL/min
- Column temp.
- 30°C
- Injection volume
- 0.5 µL
- Diluent
- mobile phase
View detail