Shinwa Chemical Industries Ltd.
LC/GC Search / Application Note

Shinwa Chemical Industries
Application Search

Search result
"Dichlorvos " Filtered 4 Item.

Graph

GC

Samples
Dichlorvos (DDVP) , Methyldemeton , Methyl parathion , Parathion , EPN
Applications
Measurement of organophosphorus compounds using Shinwasorb
Column
Silicone OV-17 1% Shinwasorb-S 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
180-250℃ (8℃/min)
Carrier gas
50 mL/min N₂
Injection port temp.
260℃
Detector temp.
260℃
Detector
FPD
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Acetone , Dichlorvos , 1-chloronaphthalene
Applications
Analysis of dichlorvos using Shinwasorb
Column
Silicone DC-550 10% Shinwasorb-S 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
180℃
Carrier gas
40 mL/min N₂ (Adjust so that the retention time of dichlorvos is about 6 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Dichlorvos , Tributyl phosphate
Applications
Analysis of dichlorvos using Shinwasorb -2-
Column
Silicone OV-225 1% Shinwasorb-U 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
170℃
Carrier gas
50 mL/min N₂(Adjust so that the retention time of tributyl phosphate is about 5 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FPD
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Dichlorvos, Diphenyl in Acetone
Applications
Analysis of dichlorvos resin strip using Shinwasorb
Column
Silicone OV-11 10% Shinwasorb-U 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
170℃
Carrier gas
35 mL/min N₂(Adjust so that the retention time of dichlorvos is about 8 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FID
Injection volume
3.0 μL

View detail