Shinwa Chemical Industries Ltd.
LC/GC Search / Application Note

Shinwa Chemical Industries
Application Search

Search result
"Methyl mercaptan" Filtered 6 Item.

Graph

GC

Samples
Carbonyl Sulfide , Hydrogen Sulfide , Carbon Disulfide , Methyl Mercaptan , Ethyl Mercaptan , Dimethyl Sulfide , Benzene(Solvent) , Dimethyl Disulfide
Applications
Sulfur Compounds of Air Pollution
Column
1,2,3-TCEP 25% Shimalite 80-100 (AW-DMCS)
Column Size
3.1m×3.2mm
Column temp.
70°C
Carrier gas
N2,50ml/min
Injection port temp.
180°C
Detector temp.
180°C
Detector
FPD

View detail


Graph

GC

Samples
Hydrogen Sulfide , Sulfur Dioxide , Methyl Mercaptan , Ethyl Mercaptan , Dimethyl Sulfide , Carbon Disulfide , Benzene , Dimethyl Disulfide
Applications
Air pollution sulfur compound (1)
Column
PPE-5rings 10% Shimalite TPA 60-80
Column Size
3.1m×3.2mm
Column temp.
70°C
Carrier gas
N2,50ml/min
Injection port temp.
180°C
Detector temp.
180°C(FPD)

View detail


Graph

GC

Samples
Methane , Ethane , Hydrogen Sulfide , Cabonyl Sulfide , Propane , Isobutane , Methyl Mercaptane , n-Butane
Applications
Sulfur compounds in Lower Hydrocarbon
Column
Sunpak-S 80-100
Column Size
2.1m×3.2mm
Column temp.
100°C
Carrier gas
N2,50ml/min
Injection port temp.
180°C
Detector temp.
180°C(FPD)

View detail


Graph

GC

Samples
Carbon Disulfide , Hydrogen Sulfide , Methyl Mercaptan ,Ethyl Mercaptan , DMS , Benzene . DMDS
Applications
Air pollution sulfur compound (3)
Column
β,β-ODPN 25% Chromosorb W 60-80 (AW-DMCS-ST)
Column Size
3.1m×3.2mm
Column temp.
70°C
Carrier gas
N2,50ml/min
Detector temp.
180°C(FPD)
Injection volume
-

View detail


Graph

GC

Samples
Carbonyl sulfide , Hydrogen sulfide , Carbon disulfide , Methyl mercaptan , Ethyl mercaptan , Dimethyl sulfide , Dimethyl disulfide
Applications
Analysis of methyl mercaptan, hydrogen sulfide, dimethyl sulfide, and dimethyl disulfide using Shinwasorb
Column
β,β’-Oxydipropionitrile 12.5% Shinwasorb-S 60/80 ST
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
70℃
Carrier gas
50 mL/min N₂
Injection port temp.
180℃
Detector temp.
180℃
Detector
FPD (S)
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Carbon sulfide , Hydrogen sulfide , Carbon disulfide , Methyl mercaptan , Ethyl mercaptane , Dimethyl sulfide
Applications
Measurements of offensive odor substances using Shinwasorb -3-
Column
1,2,3-TCEP 25% Shinwasorb-S 80/100 ST
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
70℃
Carrier gas
50 mL/min N₂
Injection port temp.
180℃
Detector temp.
180℃
Detector
FPD
Injection volume
1.0 μL

View detail