Shinwa Chemical Industries Ltd.
LC/GC Search / Application Note

Shinwa Chemical Industries
Application Search

Search result
"Primidone" Filtered 2 Item.

Graph

GC

Samples
Pyridine , 2-ethyl-2-phenylmalonediamide , Stearylalcohol ,
Applications
Analysis of primidone using Shinwasorb ~Purity test (3) 2-ethyl-2-phenylmalonediamide~
Column
Silicone OV-17 1.5% Shinwasorb-S 100/120
Column Size
1.6 m x 3.2 mm I.D. Glass
Column temp.
195℃
Carrier gas
31 mL/min N2 (Adjust so that the retention time of stearylalcohol is about 10 minutes.)
Injection port temp.
250°C
Detector temp.
250°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

LC

Samples
Ethosuximide , Primidone , Phenobarbital , Phenytoin , Carbamazepin
Applications
Anticonvulsant analysis
Column
STR ODS-II
Column Size
150 mm L × 4.6 mm I.D.
Mobile phase
100 mM Phosphoric acid buffer(pH 4.0)/CH3OH/CH3CN = 4/2/1
Frow rate
1.0 mL/min
Column temp.
40°C

View detail