Search result
"Propylene" Filtered 12 Item.
GC
- Samples
- Ethyleneoxide , Propyleneoxide , Ethanol
- Applications
- Oxirane and propylene oxide in Ethanol (2)
- Column
- Flexol 8N8 25% Chromosorb W 60-80 (AW-DMCS)
- Column Size
- 2.1m×3.2mm
- Column temp.
- 50°C
- Carrier gas
- N2,50ml/min
- Injection port temp.
- 200°C
- Detector temp.
- 200°C(FID)
View detail
GC
- Samples
- Methane , Ethane , Ethylene , Propane , Propylene , iso-Butane , n-Butane , 1-Butylene , iso-Butylene , trans-2-Butylene , iso-Pentane , cis-2-Butylene , n-Pentane , 1,3-Butadiene
- Applications
- Lower Hydrocarbons C1〜C5
- Column
- Sebaconitrile 25% Shimalite 60-80 (NAW)
- Column Size
- 12m×3mm(SUS)
- Column temp.
- 40°C
- Carrier gas
- N2,40ml/min
- Injection port temp.
- 40°C
- Detector temp.
- 40°C(FID)
View detail
GC
- Samples
- Methane , Ethane , Ethylene , Propane , Propylene , iso-Butane , n-Butane , 1-Butene , iso-Butene , trans-2-Butene , cis-2-Butene , iso-Pentane , 1,3-Butadiene , n-Pentane ,
- Applications
- Liquefied petroleum gas (LPG)
- Column
- SM-6
- Column Size
- 6.0 m × 3.0 mm Stainless steel
- Column temp.
- 40°C
- Carrier gas
- He, 25 ml/min
- Injection port temp.
- 40°C
- Detector temp.
- 40°C
- Detector
- TCD
- Injection volume
- 1.0 mL
View detail
GC
- Samples
- Methane , Ethane , Ethylene , Propane , iso-Butane , Propylene , n-Butane
- Applications
- Ethane and ethylene in the exhaust gas
- Column
- Squalane 1% Activated Alumina 60-80
- Column Size
- 1.1m×3.2mm
- Column temp.
- 40°C
- Carrier gas
- N2,30ml/min
- Injection port temp.
- 40°C
- Detector temp.
- 40°C(FID)
View detail
GC
- Samples
- Propyleneglycol , Ethyleneglycol , Decyl Alcohol , 1,4-Butanediol , Diethyleneglycol , Lauryl Alcohol , Myristyl Alcohol , Triethyleneglycol , Cetyl Alcohol , Stearyl Alcohol
- Applications
- Glycol and higher alcohol
- Column
- Thermon-3000 1% Shimalite TPA 60-80
- Column Size
- 1.1m×3.2mm
- Column temp.
- 90°C→180°C(5°C/min)
- Carrier gas
- N2,60ml/min
- Injection port temp.
- 260°C
- Detector temp.
- 260°C(FID)
View detail
GC
- Samples
- Ethylene oxide , Propylene oxide
- Applications
- Measurement of ethylene oxide using Shinwasorb
- Column
- Flexol 8N8 12.5% Shinwasorb-U 60/80
- Column Size
- 2.1 m x 3.2 mm I.D. Glass
- Column temp.
- 150℃
- Carrier gas
- 40 mL/min N₂
- Injection port temp.
- 150℃
- Detector temp.
- 150℃
- Detector
- FID
- Injection volume
- 100 μL
View detail
GC
- Samples
- Mixture of 22 organic solvents
- Applications
- Measurement of organic solvents using Shinwasorb -2-
- Column
- PEG1500 12.5% Shinwasorb-U 40/60
- Column Size
- 2.1 m x 3.2 mm I.D. Glass
- Column temp.
- 60~120℃ (4℃/min)
- Carrier gas
- 40 mL/min N₂
- Injection port temp.
- 200℃
- Detector temp.
- 200℃
- Detector
- FID
- Injection volume
- 1.0 μL
View detail
GC
- Samples
- Propyleneglycol , 1,3-propanediol
- Applications
- Measurement of food additives using Sunpak-A
- Column
- Sunpak-A 50/80
- Column Size
- 2.1 m x 3.2 mm I.D. Glass
- Column temp.
- 180℃
- Carrier gas
- 50 mL/min N₂
- Injection port temp.
- 200℃
- Detector temp.
- 230℃
- Detector
- FID
- Injection volume
- 1.0 μL
View detail
GC
- Samples
- Propyleneglycol , 1,3-propanediol
- Applications
- Measurement of food additives using Sunpak-A
- Column
- Sunpak-A 50/80
- Column Size
- 2.1 m x 3.2 mm I.D. Glass
- Column temp.
- 180℃
- Carrier gas
- 50 mL/min N₂
- Injection port temp.
- 200℃
- Detector temp.
- 230℃
- Detector
- FID
- Injection volume
- 1.0 μL
View detail