"n-Pentane" Filtered 7 Item.

Graph

GC

Samples
Methane , Ethane , Ethylene , Propane , Propylene , iso-Butane , n-Butane , 1-Butylene , iso-Butylene , trans-2-Butylene , iso-Pentane , cis-2-Butylene , n-Pentane , 1,3-Butadiene
Applications
Lower Hydrocarbons C1〜C5
Column
Sebaconitrile 25% Shimalite 60-80 (NAW)
Column Size
12m×3mm(SUS)
Column temp.
40°C
Carrier gas
N2,40ml/min
Injection port temp.
40°C
Detector temp.
40°C(FID)

View detail


Graph

GC

Samples
Methane , Ethane , Ethylene , Propane , Propylene , iso-Butane , n-Butane , 1-Butene , iso-Butene , trans-2-Butene , cis-2-Butene , iso-Pentane , 1,3-Butadiene , n-Pentane ,
Applications
Liquefied petroleum gas (LPG)
Column
SM-6
Column Size
6.0 m × 3.0 mm Stainless steel
Column temp.
40°C
Carrier gas
He, 25 ml/min
Injection port temp.
40°C
Detector temp.
40°C
Detector
TCD
Injection volume
1.0 mL

View detail


Graph

GC

Samples
n-Pentane , n-Hexane , n-Heptane , n-Octane , Acrolein , Carbon Tetrachloride , Benzene , Acrylonitrile , Acetonitrile , Perchloroethylene , Toluene , Ethylbenzene , m-Xylene , o-Xylene , n-Dodecane , Styrene monomer , n-Tridecane , p-Xylene
Applications
Organic solvent in work environment (1)
Column
Thermon-1000 25% Shimalite 80-100 (AW-DMCS)
Column Size
3.1m×3.2mm
Column temp.
130°C
Carrier gas
N2,40ml/min
Injection port temp.
240°C
Detector temp.
240°C(FID)

View detail


Graph

GC

Samples
n-Pentane , n-Hexane , Benzene , iso-Octane , Toluene , Ethylbenzene , m-Xylene , o-Xylene , p-Xylene
Applications
Premium gasoline
Column
ULBON HR-1
Column Size
50m×0.25mm(0.5μm)
Column temp.
60°C→220°C(4°C/min)
Carrier gas
He,0.75ml/min
Detector temp.
250°C(FID)
Injection volume
0.5μl

View detail


Graph

GC

Samples
n-Pentane , n-Hexane , Benzene , iso-Octane , Toluene , Ethylbenzene , m-Xylene , o-Xylene , p-Xylene
Applications
Regular gasoline
Column
ULBON HR-1
Column Size
50m×0.25mm(0.5μm)
Column temp.
60°C→220°C(4°C/min)
Carrier gas
He,0.75ml/min
Detector temp.
250°C(FID)
Injection volume
0.5μl

View detail


Graph

GC

Samples
liquefied petroleum gas (LPG)
Applications
Measurement of liquefied petroleum gas (LPG) using Shinwasorb
Column
DBM+ODPN (95:5) 35% Shinwasorb-U 60/80
Column Size
6.0 m (3.0 m+3.0 m) x 3.0 mm I.D. stainless steel
Column temp.
40℃
Carrier gas
45 mL/min N₂
Injection port temp.
50℃
Detector temp.
50℃
Detector
TCD
Injection volume
1.0 mL

View detail


Graph

GC

Samples
liquefied petroleum gas (LPG)
Applications
Measurement of liquefied petroleum gas (LPG) using Supak-A
Column
Supak-A 50/80
Column Size
2.1 m x 3.2 mm I.D. glass
Column temp.
60℃~120℃ (5°C/min.)
Carrier gas
50 mL/min N₂
Injection port temp.
200℃
Detector temp.
200℃
Detector
FID
Injection volume
0.5 mL

View detail