"OV-17" Filtered 10 Item.

Graph

GC

Samples
Ethyl Acetate , Acetaldehyde-2,4-Dinitrophenylhydrazone
Applications
Acetaldehyde -2,4- dinitrophenylhydrazone
Column
Silicone OV-17 1% Chromosorb W 80-100 (AW-DMCS)
Column Size
2.1m×3.2mm
Column temp.
220°C
Carrier gas
N2,60ml/min
Injection port temp.
270°C
Detector temp.
270°C(FID)

View detail


Graph

GC

Samples
Estrone , Estriol
Applications
Estrone and Estriol
Column
Silicone OV-17 1.5% Shimalite W 80-100 (AW-DMCS)
Column Size
1.6m×3.2mm
Column temp.
260°C
Carrier gas
N2,50ml/min
Injection port temp.
290°C
Detector temp.
290°C(FID)

View detail


Graph

GC

Samples
α-BHC , γ-BHC , β-BHC , δ-BHC , Parathion , PP´-DDE , Endrin , PP´-DDD , PP´-DDT , EPN
Applications
Chlorinated pesticide in green tea
Column
Silicone OV-17 2% Chromosorb W 80-100 (AW-DMCS)
Column Size
2.1m×3.2mm
Column temp.
200°C
Carrier gas
N2,60ml/min
Injection port temp.
270°C
Detector temp.
270°C(ECD)

View detail


Graph

GC

Samples
PCBs
Applications
Analysis of PCBs using Shinwasorb
Column
Silicone OV-17 2% Shinwasorb-U 60/80  
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
210℃
Carrier gas
50 mL/min N₂
Injection port temp.
260℃
Detector temp.
260℃
Detector
ECD
Injection volume
4.0 μL

View detail


Graph

GC

Samples
Dichlorvos (DDVP) , Methyldemeton , Methyl parathion , Parathion , EPN
Applications
Measurement of organophosphorus compounds using Shinwasorb
Column
Silicone OV-17 1% Shinwasorb-S 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
180-250℃ (8℃/min)
Carrier gas
50 mL/min N₂
Injection port temp.
260℃
Detector temp.
260℃
Detector
FPD
Injection volume
1.0 μL

View detail


Graph

GC

Samples
homatropine hydrobromide , atropine
Applications
Analysis of opium alkaloid atropine using Shinwasorb
Column
Silicone OV-17 1% Shinwasorb-S 60/80
Column Size
1.6m×3.2mm glass
Column temp.
210°C
Carrier gas
40 mL/min N2
Injection port temp.
270°C
Detector temp.
270°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
homatropine hydrobromide , scopolamine
Applications
Analysis of opium alkaloid scopolamine using Shinwasorb
Column
Silicone OV-17 1% Shinwasorb-S 60/80
Column Size
1.6m×3.2mm glass
Column temp.
210°C
Carrier gas
45 mL/min N2
Injection port temp.
250°C
Detector temp.
250°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Cyclohexane , N,N-dimethylaniline , Naphthalene ,
Applications
Analysis of ampicillin hydrate using Shinwasorb
Column
Silicone OV-17 1.5% Shinwasorb-S 60/80
Column Size
2.1 m x 2.6 mm I.D. Glass
Column temp.
105℃
Carrier gas
20 mL/min He(adjust so that the retention time of N,N-dimethylaniline is about 5 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Pyridine , 2-ethyl-2-phenylmalonediamide , Stearylalcohol ,
Applications
Analysis of primidone using Shinwasorb ~Purity test (3) 2-ethyl-2-phenylmalonediamide~
Column
Silicone OV-17 1.5% Shinwasorb-S 100/120
Column Size
1.6 m x 3.2 mm I.D. Glass
Column temp.
195℃
Carrier gas
31 mL/min N2 (Adjust so that the retention time of stearylalcohol is about 10 minutes.)
Injection port temp.
250°C
Detector temp.
250°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Acetone , Allethrin , Cinerin I , Jasmolin I , Pyrethrin I , Cinerin II , Jasmolin II , Pyrethrin II
Applications
Analysis of pyrethrum extract using Shinwasorb
Column
Silicone OV-17 1% Shinwasorb-S 80/100
Column Size
1.1 m x 3.2 mm I.D. Glass
Column temp.
205℃
Carrier gas
35 mL/min N₂(Adjust so that the retention time of Allethrin is about 3 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FPD
Injection volume
1.0 μL

View detail