Search result
"FFAP" Filtered 5 Item.
GC
- Samples
- Benzene , Toluene , Ethylbenzene , Isopropylbenzene , n-Propylbenzene , Styrene Monomer , Cyclopentanol
- Applications
- Residual styrene monomer in the packaging material
- Column
- FFAP 20% Chromosorb W 80-100 AW
- Column Size
- 3.1m×3.2mm
- Column temp.
- 130°C
- Carrier gas
- N2,50ml/min
- Injection port temp.
- 240°C
- Detector temp.
- 240°C(FID)
View detail
GC
- Samples
- Acetic Acid , Propionic Acid , Isobutyric Acid , Butyric Acid , Isovaleric Acid , Valeric Acid , Isocaproic Acid , Caproic Acid
- Applications
- Free lower fatty acids (4) (each 100ppm)
- Column
- FFAP+H3PO4(0.3%+0.3%) Graphitecarbon 60〜80
- Column Size
- 1.6m×3.2mm
- Column temp.
- 80°C→220°C(16°C/min)
- Carrier gas
- N2,60ml/min
- Injection port temp.
- 260°C
- Detector temp.
- 260°C(FID)
View detail
GC
- Samples
- Benzene , Toluene , Ethylbenzene , Isopropylbenzene , n-Propylbenzene , Styrene monomer , Cyclopentanol , N, N-Dimethylformamide ,
- Applications
- Measurement of styrene monomer in DMF using Shinwasorb
- Column
- FFAP 10% Shinwasorb-U 60/80
- Column Size
- 3.1m×3.2mm glass
- Column temp.
- 120°C
- Carrier gas
- 50 mL/min N2
- Injection port temp.
- 240°C
- Detector temp.
- 240°C
- Detector
- FID
- Injection volume
- 1.0 μL
View detail
GC
- Samples
- Ethyl acetate , di-n-butyl phthalate , Teprenone (mono-cis isomer) , Teprenone (all-trans isomer) , ,
- Applications
- Analysis of teprenone using Shinwasorb -1-
- Column
- FFAP 5% Shinwasorb-U 80/100
- Column Size
- 2.1m×3.2mm glass
- Column temp.
- 235°C
- Carrier gas
- 47 mL/min N2
- Injection port temp.
- 270°C
- Detector temp.
- 270°C
- Detector
- FID
- Injection volume
- 2.0 μL
View detail
GC
- Samples
- Ethyl acetate , di-n-butyl phthalate , Teprenone (mono-cis isomer) , Teprenone (all-trans isomer) , ,
- Applications
- Analysis of teprenone using Shinwasorb -2-
- Column
- FFAP 5% Shinwasorb-U 80/100
- Column Size
- 2.1m×3.2mm glass
- Column temp.
- 235°C
- Carrier gas
- 47 mL/min N2
- Injection port temp.
- 270°C
- Detector temp.
- 270°C
- Detector
- FID
- Injection volume
- 2.0 µL
View detail