Search result
"n-Butyl acetate" Filtered 14 Item.
GC
- Samples
- n-Hexane , Methyl Alcohol , Acetone , Isopropyl Alcohol , Ethyl Acetate , Methyl Ethyl Ketone , Benzene , Trichloroethylene , Isobutyl Alcohol , Isobutyl Acetate , Toluene , n-Butyl Acetate , Isoamyl Alcohol , Ethylbenzene , m-Xylene , o-Xylene , n-Propylbenzene , Cyclohexanone
- Applications
- Organic solvent in the air (ppm level)
- Column
- SBS-100 10% Shimalite TPA 60-80
- Column Size
- 3.1m×3.2mm
- Column temp.
- 110°C
- Carrier gas
- N2,30ml/min
- Injection port temp.
- 200°C
- Detector temp.
- 200°C(FID)
View detail
GC
- Samples
- Acetone , Isopropyl Alcohol , n-Hexane , Ethyl Acetate , Chloroform , Carbon Tetrachloride , n-Butyl Alcohol , Methyl Isobutyl Ketone , Isobutyl Acetate , Toluene , n-Butyl Acetate , Ethylbenzene , m-Xylene , o-Xylene , Cyclohexanone , Butyl Cellosolve , 1,1,2,2-Tetrachloroethane , o-Dichlorobenzene
- Applications
- Organic solvent in work environment (2)
- Column
- SBS-200 20% Shimalite W 100-120 (AW-DMCS)
- Column Size
- 3.1m×3.2mm
- Column temp.
- 100°C
- Carrier gas
- N2,50ml/min
- Injection port temp.
- 240°C
- Detector temp.
- 240°C(FID)
View detail
GC
- Samples
- Acetone , Isopropyl Alcohol , n-Hexane , Ethyl Acetate , Chloroform , Carbon Tetrachloride , n-Butyl Alcohol , Methyl Isobutyl Ketone , Isobutyl Acetate , Toluene , n-Butyl Acetate , Ethylbenzene , m-Xylene , o-Xylene , Cyclohexanone , Butyl Cellosolve , 1,1,2,2-Tetrachloroethane , o-Dichlorobenzene
- Applications
- Organic solvents of the atmosphere (ppm level)
- Column
- SBS-200 20% Shimalite W 100-120 (AW-DMCS)
- Column Size
- 3.1m×3.2mm
- Column temp.
- 100°C
- Carrier gas
- N2,50ml/min
- Injection port temp.
- 240°C
- Detector temp.
- 240°C(FID)
View detail
GC
- Samples
- Solvents
- Applications
- Solvents (5)
- Column
- ULBON HR-1
- Column Size
- 50m×0.25mm(0.5μm)
- Column temp.
- 60°C→120°C(2°C/min)
- Carrier gas
- He,0.93ml/min
- Detector temp.
- 150°C(FID)
- Injection volume
- 1μl
View detail
GC
- Samples
- Methanol , Ethyl Ether , Isopropanol , trans-1,2-Dichloroethylene , Ethyl Acetate , Chloroform , Isobutanol , Trichloroethylene , n-Butanol , 1,4-Dioxane , Toluene , Tetrachloroethylene , n-Butyl Acetate , Ethylbenzene , p-Xylene , Isoamyl Acetate , o-Xylene , n-Amyl Acetate , Cyclohexanol , Cyclohexanone , 2-Methylcyclohexanol , 2-Methylcyclohexanone , p-Dichlorobenzene , o-Dichlorobenzene , o-Cresol , m-Cresol
- Applications
- The work environment (2)
- Column
- ULBON HR-1701
- Column Size
- 30m×0.53mm(5μm)
- Column temp.
- 60°C→180°C(4°C/min)
- Carrier gas
- He,10ml/min
- Detector temp.
- 200°C(FID)
- Injection volume
- 1ml
View detail
GC
- Samples
- Methanol , Ethanol , Acetone , IPA , n-Propanol , Ethyl Acetate , MEK , Chloroform , Benzene , n-Butanol , Toluene , MIBK , Isobutyl Acetate , n-Butyl Acetate , Ethylbenzene , m-Xylene , o-Xylene , Cellosolve Acetate
- Applications
- Solvents (2)
- Column
- ULBON HR-1701
- Column Size
- 25m×0.32mm(0.5μm)
- Column temp.
- 40°C→90°C(2°C/min)
- Carrier gas
- He,0.83ml/min
- Detector temp.
- 200°C(FID)
- Injection volume
- 1μl
View detail
GC
- Samples
- Solvents (4)
- Applications
- Solvents (4)
- Column
- ULBON HR-1701
- Column Size
- 50m×0.25mm(0.5μm)
- Column temp.
- 60°C→180°C(4°C/min)
- Carrier gas
- He,1.4ml/min
- Detector temp.
- 200°C(FID)
- Injection volume
- 1μl
View detail
GC
- Samples
- n-Butyl Formate , n-Butyl Acetate , n-Butyl Propionate , n-Butyl iso-Butyrate , n-Butyl n-Butyrate , n-Butyl Crotonate , n-Butyl n-Caproate , n-Butyl Lactate , n-Butyl n-Caprate , Di-n-Butyl Oxalate , n-Butyl Benzoate , Di-n-Butyl Malonate , Di-n-Butyl Fumarate , Di-n-Butyl Succinate , Di-n-Butyl Maleate , Di-n-Butyl Adipate , Di-n-Butyl Maleate , Di-n-Butyl Phthalate , Di-n-Butyl Sebacate , Di-n-Butyl Isophthalate , Di-n-Butyl Terephthalate , Tri-n-Butyl Citrate
- Applications
- Organic acid butyl ester
- Column
- ULBON HR-20M
- Column Size
- 25m×0.25mm(0.25μm)
- Column temp.
- 80°C→240°C(4°C/min)
- Carrier gas
- He,1.1ml/min
- Detector temp.
- 250°C(FID)
- Injection volume
- 1.0μm
View detail
GC
- Samples
- Methanol , Ethanol , Acetone , n-Propanol , n-Hexane , Ethyl Acetate , Chloroform , Benzene , MIBK , Toluene , Isobutyl Acetate , n-Butyl Acetate , Ethylbenzene , m-Xylene , o-Xylene , Cellosolve Acetate
- Applications
- Solvents (1)
- Column
- ULBON HR-52
- Column Size
- 25m×0.32mm(0.5μm)
- Column temp.
- 40°C→90°C(2°C/min)
- Carrier gas
- He,0.83ml/min
- Detector temp.
- 200°C(FID)
- Injection volume
- 1μl
View detail
GC
- Samples
- Methanol , Acetone , Dichloromethane , trans-1,2-Dichloroethylene , n-Hexane , Ethyl Acetate , Tetrahydrofuran , 1,1,1-Trichloroethane , Carbon Tetrachloride , Trichloroethylene , 1,4-Dioxane , Methyl Isobutyl Ketone , Toluene , Methyl n-Butyl Ketone , n-Butyl Acetate , Tetrachloroethylene , Chlorobenzene , Ethylbenzene , p-Xylene , Styrene , o-Xylene , Butyl Cellosolve , 1,1,2,2-Tetrachloroethane , 3-Methylcyclohexanol(cis, trans) , 2-Methylcyclohexanone , 4-Methylcyclohexanone , m-Dichlorobenzene , p-Dichlorobenzene , o-Dichlorobenzene , o-Cresol , p-Cresol
- Applications
- Solvents (3)
- Column
- ULBON HR-52
- Column Size
- 50m×0.25mm(0.5μm)
- Column temp.
- 60°C→130°C(3°C/min)
- Carrier gas
- He,1.2ml/min
- Detector temp.
- 200°C(FID)
- Injection volume
- 1μl
View detail
GC
- Samples
- Ethyl acetate , di-n-butyl phthalate , Teprenone (mono-cis isomer) , Teprenone (all-trans isomer) , ,
- Applications
- Analysis of teprenone using Shinwasorb -1-
- Column
- FFAP 5% Shinwasorb-U 80/100
- Column Size
- 2.1m×3.2mm glass
- Column temp.
- 235°C
- Carrier gas
- 47 mL/min N2
- Injection port temp.
- 270°C
- Detector temp.
- 270°C
- Detector
- FID
- Injection volume
- 2.0 μL
View detail
GC
- Samples
- Ethyl acetate , di-n-butyl phthalate , Teprenone (mono-cis isomer) , Teprenone (all-trans isomer) , ,
- Applications
- Analysis of teprenone using Shinwasorb -2-
- Column
- FFAP 5% Shinwasorb-U 80/100
- Column Size
- 2.1m×3.2mm glass
- Column temp.
- 235°C
- Carrier gas
- 47 mL/min N2
- Injection port temp.
- 270°C
- Detector temp.
- 270°C
- Detector
- FID
- Injection volume
- 2.0 µL
View detail
GC
- Samples
- Diethyl ether , n-Hexane , Methanol , Methyl acetate , 2-Propanol , Ethyl acetate , Tetrahydrofuran , 2-Butanol , iso-Butanol , Toluene , n-Butyl acetate , Methyl n-butyl ketone , iso-Amyl alcohol , Ethylbenzene , m-Xylene , p-Xylene , o-Xylene , Styrene monomer ,
- Applications
- Measurement of organic solvents using SBS-120
- Column
- SBS-120 12% Shincarbon-A2 80/100
- Column Size
- 3.1 m x 3.2 mm I.D. Glass
- Column temp.
- 90℃
- Carrier gas
- 40 mL/min N₂
- Injection port temp.
- 230℃
- Detector temp.
- 230℃
- Detector
- FID
- Injection volume
- 1.0 μL
View detail