Shinwa Chemical Industries Ltd.
LC/GC Search / Application Note

Shinwa Chemical Industries
Application Search

Search result
"acetone" Filtered 36 Item.

Graph

GC

Samples
Acetic Acid , Propionic Acid , Isobutyric Acid , Butyric Acid , Isovaleric Acid , Valeric Acid , Isocaproic Acid , Caproic Acid
Applications
Free lower fatty acids in Acetone(each 200ppm)
Column
FON 20% Chromosorb W 80-100 (AW-DMCS)
Column Size
2.1m×3.2mm
Column temp.
170°C
Carrier gas
N2,50ml/min
Injection port temp.
250°C
Detector temp.
260°C(FID)

View detail


Graph

GC

Samples
n-Hexane , Methyl Alcohol , Acetone , Isopropyl Alcohol , Ethyl Acetate , Methyl Ethyl Ketone , Benzene , Trichloroethylene , Isobutyl Alcohol , Isobutyl Acetate , Toluene , n-Butyl Acetate , Isoamyl Alcohol , Ethylbenzene , m-Xylene , o-Xylene , n-Propylbenzene , Cyclohexanone
Applications
Organic solvent in the air (ppm level)
Column
SBS-100 10% Shimalite TPA 60-80
Column Size
3.1m×3.2mm
Column temp.
110°C
Carrier gas
N2,30ml/min
Injection port temp.
200°C
Detector temp.
200°C(FID)

View detail


Graph

GC

Samples
Acetone , Isopropyl Alcohol , n-Hexane , Ethyl Acetate , Chloroform , Carbon Tetrachloride , n-Butyl Alcohol , Methyl Isobutyl Ketone , Isobutyl Acetate , Toluene , n-Butyl Acetate , Ethylbenzene , m-Xylene , o-Xylene , Cyclohexanone , Butyl Cellosolve , 1,1,2,2-Tetrachloroethane , o-Dichlorobenzene
Applications
Organic solvent in work environment (2)
Column
SBS-200 20% Shimalite W 100-120 (AW-DMCS)
Column Size
3.1m×3.2mm
Column temp.
100°C
Carrier gas
N2,50ml/min
Injection port temp.
240°C
Detector temp.
240°C(FID)

View detail


Graph

GC

Samples
Acetone , Isopropyl Alcohol , n-Hexane , Ethyl Acetate , Chloroform , Carbon Tetrachloride , n-Butyl Alcohol , Methyl Isobutyl Ketone , Isobutyl Acetate , Toluene , n-Butyl Acetate , Ethylbenzene , m-Xylene , o-Xylene , Cyclohexanone , Butyl Cellosolve , 1,1,2,2-Tetrachloroethane , o-Dichlorobenzene
Applications
Organic solvents of the atmosphere (ppm level)
Column
SBS-200 20% Shimalite W 100-120 (AW-DMCS)
Column Size
3.1m×3.2mm
Column temp.
100°C
Carrier gas
N2,50ml/min
Injection port temp.
240°C
Detector temp.
240°C(FID)

View detail


Graph

GC

Samples
Benzene , Toluene , Acetone , iso-Propanol , Ethanol , Isobutanol , Methanol
Applications
The impurity component in the carbinol
Column
Sorbitol 30% Shimalite 60-80 (NAW)
Column Size
6.1m×3.2mm
Column temp.
110°C
Carrier gas
N2,50ml/min
Injection port temp.
240°C
Detector temp.
240°C(FID)

View detail


Graph

GC

Samples
Solvents
Applications
Solvents (5)
Column
ULBON HR-1
Column Size
50m×0.25mm(0.5μm)
Column temp.
60°C→120°C(2°C/min)
Carrier gas
He,0.93ml/min
Detector temp.
150°C(FID)
Injection volume
1μl

View detail


Graph

GC

Samples
Methanol , Ethanol , Acetone , IPA , n-Propanol , Ethyl Acetate , MEK , Chloroform , Benzene , n-Butanol , Toluene , MIBK , Isobutyl Acetate , n-Butyl Acetate , Ethylbenzene , m-Xylene , o-Xylene , Cellosolve Acetate
Applications
Solvents (2)
Column
ULBON HR-1701
Column Size
25m×0.32mm(0.5μm)
Column temp.
40°C→90°C(2°C/min)
Carrier gas
He,0.83ml/min
Detector temp.
200°C(FID)
Injection volume
1μl

View detail


Graph

GC

Samples
n-Hexane , Ethyl Ether , Acetone , Methyl Acetate , trans-1,2-Dichloroethylene , Methanol , Isopropyl Acetate , Isopropanol , Dichloromethane , n-Propyl Acetate , cis-1,2-Dichloroethylene , Methyl Isobutyl Ketone , Chloroform , Toluene , 1,4-Dioxane , Isobutanol , Isoamyl Acetate , Ethylbenzene , p-Xylene , m-Xylene , Methyl Cellosolve , o-Xylene , Ethyl Cellosolve , Styrene , Cellosolve Acetate , N,N-Dimethylformamide , 4-Methylcyclohexanone , Butyl Cellosolve , cis-4-Methylcyclohexanol , m-Dichlorobenzene , p-Dichlorobenzene , trans-4-Methylcyclohexanol , o-Dichlorobenzene , o-Cresol , p-Cresol , m-Cresol
Applications
Solvents (6)
Column
ULBON HR-20M
Column Size
50m×0.25mm(0.5μm)
Column temp.
60°C→240°C(3°C/min)
Carrier gas
He,1.3ml/min
Detector temp.
250°C(FID)
Injection volume
1μl

View detail


Graph

GC

Samples
Methanol , Acetone , Dichloromethane , trans-1,2-Dichloroethylene , n-Hexane , Ethyl Acetate , Tetrahydrofuran , 1,1,1-Trichloroethane , Carbon Tetrachloride , Trichloroethylene , Methyl Isobutyl Ketone , Toluene , Methyl Butyl Ketone , Tetrachloroethylene , Chlorobenzene , Ethylbenzene , p-Xylene , Styrene , 1,1,2,2-Tetrachloroethane , 2-Methylcyclohexanone , 4-Methylcyclohexanone , m-Dichlorobenzene , p-Dichlorobenzene , o-Cresol , p-Cresol
Applications
The work environment (1)
Column
ULBON HR-52
Column Size
30m×0.53mm(5μm)
Column temp.
60°C→130°C(3°C/min)
Carrier gas
He,8.0ml/min
Detector temp.
150°C(FID)
Injection volume
1ml

View detail


Graph

GC

Samples
Methanol , Ethanol , Acetone , n-Propanol , n-Hexane , Ethyl Acetate , Chloroform , Benzene , MIBK , Toluene , Isobutyl Acetate , n-Butyl Acetate , Ethylbenzene , m-Xylene , o-Xylene , Cellosolve Acetate
Applications
Solvents (1)
Column
ULBON HR-52
Column Size
25m×0.32mm(0.5μm)
Column temp.
40°C→90°C(2°C/min)
Carrier gas
He,0.83ml/min
Detector temp.
200°C(FID)
Injection volume
1μl

View detail


Graph

GC

Samples
Methanol , Acetone , Dichloromethane , trans-1,2-Dichloroethylene , n-Hexane , Ethyl Acetate , Tetrahydrofuran , 1,1,1-Trichloroethane , Carbon Tetrachloride , Trichloroethylene , 1,4-Dioxane , Methyl Isobutyl Ketone , Toluene , Methyl n-Butyl Ketone , n-Butyl Acetate , Tetrachloroethylene , Chlorobenzene , Ethylbenzene , p-Xylene , Styrene , o-Xylene , Butyl Cellosolve , 1,1,2,2-Tetrachloroethane , 3-Methylcyclohexanol(cis, trans) , 2-Methylcyclohexanone , 4-Methylcyclohexanone , m-Dichlorobenzene , p-Dichlorobenzene , o-Dichlorobenzene , o-Cresol , p-Cresol
Applications
Solvents (3)
Column
ULBON HR-52
Column Size
50m×0.25mm(0.5μm)
Column temp.
60°C→130°C(3°C/min)
Carrier gas
He,1.2ml/min
Detector temp.
200°C(FID)
Injection volume
1μl

View detail


Graph

GC

Samples
Dichlorvos (DDVP) , Methyldemeton , Methyl parathion , Parathion , EPN
Applications
Measurement of organophosphorus compounds using Shinwasorb
Column
Silicone OV-17 1% Shinwasorb-S 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
180-250℃ (8℃/min)
Carrier gas
50 mL/min N₂
Injection port temp.
260℃
Detector temp.
260℃
Detector
FPD
Injection volume
1.0 μL

View detail


Graph

GC

Samples
2,6-Xylenol , o-Cresol , Phenol , 2,4-Xylenol , p-Cresol , m-Cresol , 3,5-Xylenol , 3,4-Xylenol
Applications
Measurements of cresol and xylenol isomers using Shinwasorb
Column
FAP-S Shinwasorb-U 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
160℃
Carrier gas
50 mL/min N₂
Injection port temp.
250℃
Detector temp.
250℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Acetic acid , Propionic acid , Isobutyric acid , Butyric acid , Isovaleric acid , Valeric acid , Isocaproic acid , Capric acid , Enanthic acid
Applications
Measurement of short-chain free fatty acids using Shinwasorb -1-
Column
Thermon-1000+H3PO4 2.5+0.25% Shinwasorb-U 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
150℃
Carrier gas
50 mL/min N₂
Injection port temp.
250℃
Detector temp.
240℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Acetic acid , Propionic acid , Isobutyric acid , Butyric acid , Isovaleric acid , Valeric acid , Isocaproic acid , Capric acid , Enanthic acid
Applications
Measurement of short-chain free fatty acids using Shinwasorb -2-
Column
FAL-M Shinwasorb-S 60/80 PA
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
150℃
Carrier gas
50 mL/min N₂
Injection port temp.
260℃
Detector temp.
260℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Mixture of 13 organic solvents
Applications
Measurement of organic solvents using Shinwasorb -1-
Column
PEG1500 12.5% Shinwasorb-U 40/60
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
70℃
Carrier gas
40 mL/min N₂
Injection port temp.
200℃
Detector temp.
200℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Mixture of 22 organic solvents
Applications
Measurement of organic solvents using Shinwasorb -2-
Column
PEG1500 12.5% Shinwasorb-U 40/60
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
60~120℃ (4℃/min)
Carrier gas
40 mL/min N₂
Injection port temp.
200℃
Detector temp.
200℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Sorbic acid , Dehydroacetic acid , Benzoic acid , trans-Stilbene
Applications
Analysis of preservatives using Shinwasorb -1-
Column
DEGS+H3PO4 5.0+0.5% Shinwasorb-S 40/60
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
185℃
Carrier gas
50 mL/min N₂
Injection port temp.
260℃
Detector temp.
260℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
preservatives
Applications
Analysis of preservatives using Shinwasorb -2-
Column
Advance-DS+H3PO4 2+0.25% Shinwasorb-S 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
200℃
Carrier gas
50 mL/min N₂
Injection port temp.
260℃
Detector temp.
260℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
acetone , styrene , , ,
Applications
Analysis of polystyrene sulfonate calcium using Shinwasorb
Column
PEG20M 7.5% Shinwasorb-S 80/100
Column Size
2.1m×3.2mm glass
Column temp.
90°C
Carrier gas
60 mL/min N2
Injection port temp.
260°C
Detector temp.
260°C
Detector
FID
Injection volume
5.0 µL

View detail


Graph

GC

Samples
Acetone , Ethyl acetate
Applications
Analysis of piperacillin hydrate using Sunpak-H ~(4) Residual solvents~
Column
Sunpak-H 80/100
Column Size
1.1m x 3.2 mm I.D. Glass
Column temp.
145℃
Carrier gas
33 mL/min N2 (Adjust so that the retention time of ethyl acetate is about 4 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
o-Dichlorobenzene, m-Bromochlorobenzene in acetone
Applications
Analysis of o-dichlorobenzene using Shinwasorb
Column
PEG20M-TPA 2.5% Shinwasorb-U 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
115℃
Carrier gas
39 mL/min N₂ (Adjust so that the retention time of o-dichlorobenzene is about 11 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Acetone , cis-Allethrin , trans-Allethrin
Applications
Analysis of allethrin using Shinwasorb Part 2 ~Isomer ratio~
Column
Silicone DC QF-1 2.5% Shinwasorb-U 80/100
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
165℃
Carrier gas
45 mL/min N₂ (Adjust so that the retention time of trans-allethrin is about 40 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Acetone , Allethrin , Di(2-ethylhexyl) phthalate
Applications
Analysis of allethrin using Shinwasorb Part 3
Column
Diethyleneglycol Succinate (DEGS) 2.5% Shinwasorb-U 80/100
Column Size
1.1 m x 3.2 mm I.D. Glass
Column temp.
200℃
Carrier gas
36 mL/min N₂ (Adjust so that the retention time of allethrin is about 8 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Acetone , Dichlorvos , 1-chloronaphthalene
Applications
Analysis of dichlorvos using Shinwasorb
Column
Silicone DC-550 10% Shinwasorb-S 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
180℃
Carrier gas
40 mL/min N₂ (Adjust so that the retention time of dichlorvos is about 6 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Acetone , Allethrin , Cinerin I , Jasmolin I , Pyrethrin I , Cinerin II , Jasmolin II , Pyrethrin II
Applications
Analysis of pyrethrum extract using Shinwasorb
Column
Silicone OV-17 1% Shinwasorb-S 80/100
Column Size
1.1 m x 3.2 mm I.D. Glass
Column temp.
205℃
Carrier gas
35 mL/min N₂(Adjust so that the retention time of Allethrin is about 3 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FPD
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Dichlorvos, Diphenyl in Acetone
Applications
Analysis of dichlorvos resin strip using Shinwasorb
Column
Silicone OV-11 10% Shinwasorb-U 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
170℃
Carrier gas
35 mL/min N₂(Adjust so that the retention time of dichlorvos is about 8 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FID
Injection volume
3.0 μL

View detail


Graph

GC

Samples
Diazinon, Di-n-butyl phthalate in Acetone
Applications
Analysis of diazinon using Shinwasorb
Column
Silicone OV-11 10% Shinwasorb-U 60/80
Column Size
1.6 m x 3.2 mm I.D. Glass
Column temp.
200℃
Carrier gas
25 mL/min N₂(Adjust so that the retention time of diazinon is about 7 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FID
Injection volume
3.0 μL

View detail


Graph

GC

Samples
Anhydrous caffeine, Fenitrothion in Acetone
Applications
Analysis of Fenitrothion Emulsifiable Concentrate using Shinwasorb
Column
Silicone OV-210 3% Shinwasorb-S 80/100
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
175℃
Carrier gas
51 mL/min N₂(Adjust so that the retention time of Anhydrous caffeine is about 10 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Anhydrous caffeine, Fenitrothion in Acetone
Applications
Analysis of Fenitrothion Microcapsule using Shinwasorb
Column
Silicone OV-210 1.5% Shinwasorb-U 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
165℃
Carrier gas
50 mL/min N₂(Adjust so that the retention time of diazinon is about 17 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Benzyl salicylate, Fenitrothion in Acetone
Applications
Analysis of fenitrothion, phthalthrin mixed emulsifiable concentrate using Shinwasorb -1-
Column
Silicone OV-210 1.5% Shinwasorb-U 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
175℃
Carrier gas
52 mL/min N₂ (Adjust so that the retention time of fenitrothion is about 15 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Phthalthrin in Acetone
Applications
Analysis of phthalthrin using Shinwasorb ~Isomer ratio phthalthrin~
Column
Silicone DC QF-1 2.5% Shinwasorb-S 80/100
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
200℃
Carrier gas
40 mL/min N₂( Adjusted so that the retention time of 3.Phthalthrin is about 40 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Organic solvents
Applications
Analysis of organic solvents using Sunpak-A -1-
Column
Sunpak-A 50/80
Column Size
2.0 m x 3.0 mm I.D. Glass
Column temp.
150℃ ~ 220℃ (3 min hold) program rate 4℃/min
Carrier gas
50 mL/min N₂
Injection port temp.
250°C
Detector temp.
250°C
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Organic solvents
Applications
Analysis of organic solvents using Sunpak-A -2-
Column
Sunpak-A 50/80
Column Size
2.0 m x 3.0 mm I.D. Glass
Column temp.
150℃ ~ 220℃ (3 min hold) program rate 4℃/min
Carrier gas
50 mL/min N₂
Injection port temp.
250°C
Detector temp.
250°C
Detector
FID
Injection volume
1.0 μL

View detail


Graph

LC

Samples
Acetylacetone
Applications
The elution behavior of the sample which forms metal complex
Column
STR ODS-II
Column Size
150 mm L × 4.6 mm I.D.
Mobile phase
20 mM Phosphoric acid buffer (pH 7.6)/CH3OH = 30/70
Frow rate
1.0 mL/min
Column temp.
40°C

View detail


Graph

LC

Samples
Formaldehyde , Methanol , Acetone
Applications
Formaldehyde , Methanol , Acetone
Column
ULTRON PS-80H
Column Size
300 mm L × 8.0 mm I.D.
Mobile phase
HClO4 aq
Frow rate
1.0 mL/min
Column temp.
60°C
Injection volume
10 µL
Diluent
mobile phase

View detail