Shinwa Chemical Industries Ltd.
LC/GC Search / Application Note

Shinwa Chemical Industries
Application Search

Search result
"Shinwasorb" Filtered 70 Item.

Graph

GC

Samples
Chloroform , Fenitrothion , Fluoranthene
Applications
Analysis of fenitrothion using Shinwasorb
Column
Polyphenyl Ether (5 rings) 1.5% Shinwasorb-S 100/120
Column Size
1.1 m x 3.2 mm I.D. Glass
Column temp.
195℃
Carrier gas
55 mL/min N₂
Injection port temp.
230℃
Detector temp.
230℃
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Carbonyl sulfide , Hydrogen sulfide , Carbon disulfide , Methyl mercaptan , Ethyl mercaptan , Dimethyl sulfide , Dimethyl disulfide
Applications
Analysis of methyl mercaptan, hydrogen sulfide, dimethyl sulfide, and dimethyl disulfide using Shinwasorb
Column
β,β’-Oxydipropionitrile 12.5% Shinwasorb-S 60/80 ST
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
70℃
Carrier gas
50 mL/min N₂
Injection port temp.
180℃
Detector temp.
180℃
Detector
FPD (S)
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Trimethylamine , Dimethylamine , Methylamine
Applications
Analysis of methylamines using Shinwasorb
Column
Diglycerol+TEP+KOH 7.5+7.5+1% Shinwasorb-S 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
70℃
Carrier gas
50 mL/min N₂
Injection port temp.
220℃
Detector temp.
220℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
PCBs
Applications
Analysis of PCBs using Shinwasorb
Column
Silicone OV-17 2% Shinwasorb-U 60/80  
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
210℃
Carrier gas
50 mL/min N₂
Injection port temp.
260℃
Detector temp.
260℃
Detector
ECD
Injection volume
4.0 μL

View detail


Graph

GC

Samples
PCBs
Applications
Analysis of PCBs using Shinwasorb
Column
Silicone OV-1 2% Shinwasorb-U 60/80  
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
195℃
Carrier gas
50 mL/min N₂
Injection port temp.
260℃
Detector temp.
260℃
Detector
ECD
Injection volume
4.0 μL

View detail


Graph

GC

Samples
Ethylene oxide , Propylene oxide
Applications
Measurement of ethylene oxide using Shinwasorb
Column
Flexol 8N8 12.5% Shinwasorb-U 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
150℃
Carrier gas
40 mL/min N₂
Injection port temp.
150℃
Detector temp.
150℃
Detector
FID
Injection volume
100 μL

View detail


Graph

GC

Samples
Dichlorvos (DDVP) , Methyldemeton , Methyl parathion , Parathion , EPN
Applications
Measurement of organophosphorus compounds using Shinwasorb
Column
Silicone OV-17 1% Shinwasorb-S 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
180-250℃ (8℃/min)
Carrier gas
50 mL/min N₂
Injection port temp.
260℃
Detector temp.
260℃
Detector
FPD
Injection volume
1.0 μL

View detail


Graph

GC

Samples
2,6-Xylenol , o-Cresol , Phenol , 2,4-Xylenol , p-Cresol , m-Cresol , 3,5-Xylenol , 3,4-Xylenol
Applications
Measurements of cresol and xylenol isomers using Shinwasorb
Column
FAP-S Shinwasorb-U 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
160℃
Carrier gas
50 mL/min N₂
Injection port temp.
250℃
Detector temp.
250℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Methyl mercuric chloride , Ethyl mercuric chloride
Applications
Measurement of alkyl mercury compounds (each 100 ppb in n-hexane) using Shinwasorb
Column
Thermon-HG 5% Shinwasorb-U 60/80
Column Size
0.5 m x 3.0 mm I.D. Glass
Column temp.
160℃
Carrier gas
50 mL/min N₂
Injection port temp.
250℃
Detector temp.
250℃
Detector
ECD
Injection volume
1.0 μL

View detail


Graph

GC

Samples
long-chain fatty acid methyl esters (FAMEs)
Applications
Measurement of long-chain fatty acid methyl esters (FAMEs C10~C24) using Shinwasorb
Column
Shinchrom E71 5% Shinwasorb-U 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
230℃
Carrier gas
40 mL/min N₂
Injection port temp.
250℃
Detector temp.
250℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Ethylbenzene , p-Xylene , m-Xylene , o-Xylene
Applications
Measurement of xylene isomers using Shinwasorb
Column
Shinchrom F51+Bentone 34 3+2% Shinwasorb-U 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
80℃
Carrier gas
40 mL/min N₂
Injection port temp.
240℃
Detector temp.
240℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Dichloromethane , trans-1,2-Dichloroethylene , Chloroform , 1,2-Dichloroethane , 1,1,1-Trichloroethane , Carbon tetrachloride , Trichloroethylene , Toluene , Tetrachloroethylene , Chlorobenzene , 1,1,2,2,-Tetrachloroethane , o-Dichlorobenzene
Applications
Measurement of chlorinated solvents in air using Shinwasorb
Column
SBS-300 10% Shinwasorb-S 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
100℃
Carrier gas
30 mL/min N₂
Injection port temp.
240℃
Detector temp.
240℃
Detector
FID
Injection volume
1.0 mL

View detail


Graph

GC

Samples
palm oil
Applications
Measurement of fatty acid methyl esters (palm oil) using Shinwasorb
Column
Shinchrom E71 5% Shinwasorb-U 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
210℃
Carrier gas
40 mL/min N₂
Injection port temp.
260℃
Detector temp.
260℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
beef tallow
Applications
Measurement of Measurement of beef tallow fatty acid methyl esters (FAMEs) using Shinwasorb
Column
Shinchrom E71 5% Shinwasorb-U 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
210℃
Carrier gas
40 mL/min N₂
Injection port temp.
260℃
Detector temp.
260℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
liquefied petroleum gas (LPG)
Applications
Measurement of liquefied petroleum gas (LPG) using Shinwasorb
Column
DBM+ODPN (95:5) 35% Shinwasorb-U 60/80
Column Size
6.0 m (3.0 m+3.0 m) x 3.0 mm I.D. stainless steel
Column temp.
40℃
Carrier gas
45 mL/min N₂
Injection port temp.
50℃
Detector temp.
50℃
Detector
TCD
Injection volume
1.0 mL

View detail


Graph

GC

Samples
Palmitic acid , Stearic acid , Oleic acid , Linoleic acid , Linolenic acid
Applications
Measurement of fatty acid methyl esters (soy bean oil) using Shinwasorb
Column
Shinchrom E71 5% Shinwasorb-U 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
230℃
Carrier gas
40 mL/min N₂
Injection port temp.
260℃
Detector temp.
260℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
1,1,1-Trichloroethane , Trichloroethylene , Bromodichloromethane , Chlorodibromomethane , Tetrachloroethylene , Bromoform
Applications
Measurements of trihalomethanes using Shinwasorb -1-
Column
Silicone DC-200 10% Shinwasorb-S 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
100℃
Carrier gas
50 mL/min N₂
Injection port temp.
240℃
Detector temp.
240℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
1,1,1-Trichloroethane , Trichloroethylene , Bromodichloromethane , Chlorodibromomethane , Tetrachloroethylene , Bromoform
Applications
Measurements of trihalomethanes using Shinwasorb -2-
Column
Silicone DC-550 10% Shinwasorb-S 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
90℃
Carrier gas
50 mL/min N₂
Injection port temp.
220℃
Detector temp.
220℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Acetic acid , Propionic acid , Isobutyric acid , Butyric acid , Isovaleric acid , Valeric acid , Isocaproic acid , Capric acid , Enanthic acid
Applications
Measurement of short-chain free fatty acids using Shinwasorb -1-
Column
Thermon-1000+H3PO4 2.5+0.25% Shinwasorb-U 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
150℃
Carrier gas
50 mL/min N₂
Injection port temp.
250℃
Detector temp.
240℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Acetic acid , Propionic acid , Isobutyric acid , Butyric acid , Isovaleric acid , Valeric acid , Isocaproic acid , Capric acid , Enanthic acid
Applications
Measurement of short-chain free fatty acids using Shinwasorb -2-
Column
FAL-M Shinwasorb-S 60/80 PA
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
150℃
Carrier gas
50 mL/min N₂
Injection port temp.
260℃
Detector temp.
260℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Benzene , Toluene , Ethylbenzene , Isopropylbenzene , n-Propylbenzene , Styrene monomer , Cyclopentanol , N,N-Dimethylformamide
Applications
Measurement of styrene monomer in DMF using Shinwasorb
Column
FON 10% Shinwasorb-S 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
110℃
Carrier gas
50 mL/min N₂
Injection port temp.
230℃
Detector temp.
230℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Methyl mercuric chloride
Applications
Measurement of methylmercury in black bass using Shinwasorb
Column
Thermon-HG 5% Shinwasorb-U 60/80
Column Size
0.5 m x 3.0 mm I.D. Glass
Column temp.
170℃
Carrier gas
50 mL/min N₂
Injection port temp.
200℃
Detector temp.
200℃
Detector
ECD
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Methyl mercuric chloride
Applications
Measurement of methylmercury in tuna using Shinwasorb
Column
Thermon-HG 5% Shinwasorb-U 60/80
Column Size
0.5 m x 3.0 mm I.D. Glass
Column temp.
170℃
Carrier gas
50 mL/min N₂
Injection port temp.
200℃
Detector temp.
200℃
Detector
ECD
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Mixture of 13 organic solvents
Applications
Measurement of organic solvents using Shinwasorb -1-
Column
PEG1500 12.5% Shinwasorb-U 40/60
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
70℃
Carrier gas
40 mL/min N₂
Injection port temp.
200℃
Detector temp.
200℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Mixture of 22 organic solvents
Applications
Measurement of organic solvents using Shinwasorb -2-
Column
PEG1500 12.5% Shinwasorb-U 40/60
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
60~120℃ (4℃/min)
Carrier gas
40 mL/min N₂
Injection port temp.
200℃
Detector temp.
200℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Sorbic acid , Dehydroacetic acid , Benzoic acid , trans-Stilbene
Applications
Analysis of preservatives using Shinwasorb -1-
Column
DEGS+H3PO4 5.0+0.5% Shinwasorb-S 40/60
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
185℃
Carrier gas
50 mL/min N₂
Injection port temp.
260℃
Detector temp.
260℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
preservatives
Applications
Analysis of preservatives using Shinwasorb -2-
Column
Advance-DS+H3PO4 2+0.25% Shinwasorb-S 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
200℃
Carrier gas
50 mL/min N₂
Injection port temp.
260℃
Detector temp.
260℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Methyl iodide , iso-Propyl iodide , n-Octane , o-Xylene
Applications
Analysis of hypromellose using Shinwasorb
Column
Silicone DC-200 10% Shinwasorb-S 100/120
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
90℃
Carrier gas
38 mL/min N₂
Injection port temp.
200℃
Detector temp.
200℃
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Carbon sulfide , Hydrogen sulfide , Carbon disulfide , Methyl mercaptan , Ethyl mercaptane , Dimethyl sulfide
Applications
Measurements of offensive odor substances using Shinwasorb -3-
Column
1,2,3-TCEP 25% Shinwasorb-S 80/100 ST
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
70℃
Carrier gas
50 mL/min N₂
Injection port temp.
180℃
Detector temp.
180℃
Detector
FPD
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Fragrance
Applications
Measurements of fragrance using Shinwasorb
Column
Thermon-3000 5% Shinwasorb-S 80/100
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
80~260℃(6℃/min)
Carrier gas
50 mL/min N₂
Injection port temp.
280℃
Detector temp.
280℃
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Chloroform , Ethanol , Naphthalene , Amantadine
Applications
Measurements of amantadine hydrochloride using Shinwasorb
Column
Apiezone L+KOH 2+1% Shinwasorb-S 80/100
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
125℃(5 min hold)~150℃(5℃/min)~150℃ (10 min hold)
Carrier gas
50 mL/min N₂
Injection port temp.
200℃
Detector temp.
200℃
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
200 ppm in hexane
Applications
Measurements of eucalyptus oil using Shinwasorb
Column
Thermon-3000 5% Shinwasorb-S 80/100
Column Size
5.1 m x 3.2 mm I.D. Glass
Column temp.
120℃ & 130℃
Carrier gas
40 mL/min N₂
Injection port temp.
200℃
Detector temp.
200℃
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
homatropine hydrobromide , atropine
Applications
Analysis of opium alkaloid atropine using Shinwasorb
Column
Silicone OV-17 1% Shinwasorb-S 60/80
Column Size
1.6m×3.2mm glass
Column temp.
210°C
Carrier gas
40 mL/min N2
Injection port temp.
270°C
Detector temp.
270°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
homatropine hydrobromide , scopolamine
Applications
Analysis of opium alkaloid scopolamine using Shinwasorb
Column
Silicone OV-17 1% Shinwasorb-S 60/80
Column Size
1.6m×3.2mm glass
Column temp.
210°C
Carrier gas
45 mL/min N2
Injection port temp.
250°C
Detector temp.
250°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Benzene , Toluene , Ethylbenzene , Isopropylbenzene , n-Propylbenzene , Styrene monomer , Cyclopentanol , N, N-Dimethylformamide ,
Applications
Measurement of styrene monomer in DMF using Shinwasorb
Column
FFAP 10% Shinwasorb-U 60/80
Column Size
3.1m×3.2mm glass
Column temp.
120°C
Carrier gas
50 mL/min N2
Injection port temp.
240°C
Detector temp.
240°C
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
decane , undecane , dodecane , tridecane , tetradecane , pentadecane , hexadecane , heptadecane , octadecane , nonadecane , eicosane , heneicosane , docosane
Applications
Analysis of mixture of kerosene and gas oil using Shinwasorb
Column
Silicone SE-52 5% Shinwasorb-U 40/60
Column Size
2.5 m x 3.0 mm I.D. Stainless steel
Column temp.
80~250℃ (6℃/min), 250℃ (12 min hold)
Carrier gas
50 mL/min N2
Injection port temp.
270°C
Detector temp.
270°C
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Ethyl acetate , di-n-butyl phthalate , Teprenone (mono-cis isomer) , Teprenone (all-trans isomer) , ,
Applications
Analysis of teprenone using Shinwasorb -1-
Column
FFAP 5% Shinwasorb-U 80/100
Column Size
2.1m×3.2mm glass
Column temp.
235°C
Carrier gas
47 mL/min N2
Injection port temp.
270°C
Detector temp.
270°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
n-valeric acid , valproic acid
Applications
Analysis of valproate sodium using Shinwasorb
Column
DEGA+H3PO4 2.5+0.5% Shinwasorb-S 80/100
Column Size
2.1m×3.2mm glass
Column temp.
145°C
Carrier gas
35 mL/min N2
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Ethanol , Dl-camphor , Methyl salicylate ,
Applications
Analysis of dl-camphor using Shinwasorb
Column
PEG 20M 5% Shinwasorb-S 60/80
Column Size
3.1m×3.2mm glass
Column temp.
160°C
Carrier gas
45 mL/min N2
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Ethyl acetate , di-n-butyl phthalate , Teprenone (mono-cis isomer) , Teprenone (all-trans isomer) , ,
Applications
Analysis of teprenone using Shinwasorb -2-
Column
FFAP 5% Shinwasorb-U 80/100
Column Size
2.1m×3.2mm glass
Column temp.
235°C
Carrier gas
47 mL/min N2
Injection port temp.
270°C
Detector temp.
270°C
Detector
FID
Injection volume
2.0 µL

View detail


Graph

GC

Samples
acetone , styrene , , ,
Applications
Analysis of polystyrene sulfonate calcium using Shinwasorb
Column
PEG20M 7.5% Shinwasorb-S 80/100
Column Size
2.1m×3.2mm glass
Column temp.
90°C
Carrier gas
60 mL/min N2
Injection port temp.
260°C
Detector temp.
260°C
Detector
FID
Injection volume
5.0 µL

View detail


Graph

GC

Samples
ethanol , n-ethyl caprylate , L-menthol , , ,
Applications
Analysis of polystyrene peppermint oil using Shinwasorb
Column
PEG6000 12.5% Shinwasorb-S 60/80
Column Size
2.1m×3.2mm glass
Column temp.
140°C
Carrier gas
33 mL/min N2
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
1.0 µL

View detail


Graph

GC

Samples
docosanoic acid methyl , ethyl icosapentate ,
Applications
Analysis of ethyl icosapentate using Shinwasorb
Column
DEGS 12.5% Shinwasorb-S 60/80
Column Size
1.8 m x 4.0 mm I.D. Glass
Column temp.
210℃
Carrier gas
47 mL/min N2
Injection port temp.
230℃
Detector temp.
200℃  
Detector
FID
Injection volume
3.0 μL

View detail


Graph

GC

Samples
N,N-dimethylformamide , Triphenylantimony , Spectinomycin ,
Applications
Analysis of spectinomycin hydrochloride hydrate using Shinwasorb
Column
Silicone SE-52 2.5% Shinwasorb-S 80/100
Column Size
0.6 m x 3.2 mm I.D. Glass
Column temp.
180℃
Carrier gas
37 mL/min N2
Injection port temp.
215℃
Detector temp.
220℃  
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Petroleum ether , Methyl oleate
Applications
Analysis of oleic acid using Shinwasorb
Column
Shinchrom E71 10% Shinwasorb-U 80/100
Column Size
2.1m×3.2mm
Column temp.
210°C
Carrier gas
N2,21 ml/min
Injection port temp.
250°C
Detector temp.
250°C
Detector
FID
Injection volume
0.2 μL

View detail


Graph

GC

Samples
Ethyl acetate , Caffeine , o-Toluene sulfonamide
Applications
Analysis of saccharin using Shinwasorb
Column
DEGS 1.5% Shinwasorb-S 60/80
Column Size
1.0 m x 3.0 mm I.D. Glass
Column temp.
210°C
Carrier gas
N2,55ml/min
Injection port temp.
225°C
Detector temp.
250°C
Detector
FID

View detail


Graph

GC

Samples
Limonene , Cineole
Applications
Analysis of eucalyptus oil using Shinwasorb
Column
Thermon-3000 5% Shinwasorb-S 80/100
Column Size
5.1 m x 3.2 mm I.D. Glass
Column temp.
130°C
Carrier gas
N2,35ml/min
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID

View detail


Graph

GC

Samples
Dichloromethane , l-menthol ,
Applications
Analysis of l-menthol using Shinwasorb
Column
PEG6000 12.5% Shinwasorb-S 60/80
Column Size
2.1m x3.2 mm I.D. Glass
Column temp.
140°C
Carrier gas
N2,33ml/min
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
5.0 μL

View detail


Graph

GC

Samples
Pyridine , N,N-dimethylformamide
Applications
Analysis of niceritrol using Shinwasorb
Column
PEG20M 5% Shinwasorb-S 80/100
Column Size
3.1m x3.2 mm I.D. Glass
Column temp.
160°C
Carrier gas
N2, 55ml/min
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Ethylene glycol , Diethylene glycol
Applications
Analysis of macrogol 400 using Shinwasorb
Column
Sorbitol 3% Shinwasorb-S 80/100
Column Size
1.6 m x 3.2 mm I.D. Glass
Column temp.
160℃
Carrier gas
47 mL/min N2
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Cyclohexane , N,N-dimethylaniline , Naphthalene ,
Applications
Analysis of ampicillin hydrate using Shinwasorb
Column
Silicone OV-17 1.5% Shinwasorb-S 60/80
Column Size
2.1 m x 2.6 mm I.D. Glass
Column temp.
105℃
Carrier gas
20 mL/min He(adjust so that the retention time of N,N-dimethylaniline is about 5 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Cyclohexane , Ethyl acetate , 2-propanol ,
Applications
Analysis of lenampicillin hydrochloride using Shinwasorb
Column
Quadrol 5% Shinwasorb-S 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
100℃
Carrier gas
70 mL/min N₂ (Adjust so that the retention time of cyclohexane is about 1 minute.)
Injection port temp.
160°C
Detector temp.
160°C
Detector
FID
Injection volume
4.0 μL

View detail


Graph

GC

Samples
Enflurane , 2-propanol ,
Applications
Analysis of enflurane using Shinwasorb
Column
DEGS 10% Shinwasorb-S 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
80℃
Carrier gas
32 mL/min He (Adjust so that the retention time of enflurane is about 3 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
5.0 μL

View detail


Graph

GC

Samples
Isoflurane , Ethyl acetate
Applications
Analysis of isoflurane using Shinwasorb
Column
Igepal CO-880+Ucon LB-550X 5+7.5% Shinwasorb-S 100/120
Column Size
3.5 m x 3.0 mm I.D. Stainless steel
Column temp.
80℃
Carrier gas
37 mL/min N2 (Adjust so that the retention time of isoflurane is about 7 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Pyridine , 2-ethyl-2-phenylmalonediamide , Stearylalcohol ,
Applications
Analysis of primidone using Shinwasorb ~Purity test (3) 2-ethyl-2-phenylmalonediamide~
Column
Silicone OV-17 1.5% Shinwasorb-S 100/120
Column Size
1.6 m x 3.2 mm I.D. Glass
Column temp.
195℃
Carrier gas
31 mL/min N2 (Adjust so that the retention time of stearylalcohol is about 10 minutes.)
Injection port temp.
250°C
Detector temp.
250°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Cyclohexane , Dicyclohexyl phthalate , Etofenprox
Applications
Analysis of etofenprox using Shinwasorb
Column
Silicone OV-225 2.5% Shinwasorb-U 60/80
Column Size
1.1m x 3.2 mm I.D. Glass
Column temp.
240℃
Carrier gas
37 mL/min N₂ (Adjust so that the retention time of dicyclohexyl phthalate is about 8 minutes.)
Injection port temp.
250°C
Detector temp.
250°C
Detector
FID
Injection volume
3.0 μL

View detail


Graph

GC

Samples
Etofenprox in Cyclohexane
Applications
Analysis of etofenprox using Shinwasorb -2- ~Purity (3) Related substances~
Column
Silicone OV-225 2.5% Shinwasorb-U 60/80
Column Size
1.1m x 3.2 mm I.D. Glass
Column temp.
240℃
Carrier gas
37 mL/min N₂
Injection port temp.
250°C
Detector temp.
250°C
Detector
FID
Injection volume
3.0 μL

View detail


Graph

GC

Samples
o-Dichlorobenzene, m-Bromochlorobenzene in acetone
Applications
Analysis of o-dichlorobenzene using Shinwasorb
Column
PEG20M-TPA 2.5% Shinwasorb-U 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
115℃
Carrier gas
39 mL/min N₂ (Adjust so that the retention time of o-dichlorobenzene is about 11 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Acetone , cis-Allethrin , trans-Allethrin
Applications
Analysis of allethrin using Shinwasorb Part 2 ~Isomer ratio~
Column
Silicone DC QF-1 2.5% Shinwasorb-U 80/100
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
165℃
Carrier gas
45 mL/min N₂ (Adjust so that the retention time of trans-allethrin is about 40 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Acetone , Allethrin , Di(2-ethylhexyl) phthalate
Applications
Analysis of allethrin using Shinwasorb Part 3
Column
Diethyleneglycol Succinate (DEGS) 2.5% Shinwasorb-U 80/100
Column Size
1.1 m x 3.2 mm I.D. Glass
Column temp.
200℃
Carrier gas
36 mL/min N₂ (Adjust so that the retention time of allethrin is about 8 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Acetone , Dichlorvos , 1-chloronaphthalene
Applications
Analysis of dichlorvos using Shinwasorb
Column
Silicone DC-550 10% Shinwasorb-S 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
180℃
Carrier gas
40 mL/min N₂ (Adjust so that the retention time of dichlorvos is about 6 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Chloroform , Fenitrothion , Fluoranthene
Applications
Analysis of fenitrothion using Shinwasorb
Column
Polyphenyl Ether (5 rings) 1.5% Shinwasorb-S 100/120
Column Size
1.1 m x 3.2 mm I.D. Glass
Column temp.
195℃
Carrier gas
55 mL/min N₂ (Adjust so that the retention time of fenitrothion is about 16 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Dichlorvos , Tributyl phosphate
Applications
Analysis of dichlorvos using Shinwasorb -2-
Column
Silicone OV-225 1% Shinwasorb-U 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
170℃
Carrier gas
50 mL/min N₂(Adjust so that the retention time of tributyl phosphate is about 5 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FPD
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Acetone , Allethrin , Cinerin I , Jasmolin I , Pyrethrin I , Cinerin II , Jasmolin II , Pyrethrin II
Applications
Analysis of pyrethrum extract using Shinwasorb
Column
Silicone OV-17 1% Shinwasorb-S 80/100
Column Size
1.1 m x 3.2 mm I.D. Glass
Column temp.
205℃
Carrier gas
35 mL/min N₂(Adjust so that the retention time of Allethrin is about 3 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FPD
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Dichlorvos, Diphenyl in Acetone
Applications
Analysis of dichlorvos resin strip using Shinwasorb
Column
Silicone OV-11 10% Shinwasorb-U 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
170℃
Carrier gas
35 mL/min N₂(Adjust so that the retention time of dichlorvos is about 8 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FID
Injection volume
3.0 μL

View detail


Graph

GC

Samples
Diazinon, Di-n-butyl phthalate in Acetone
Applications
Analysis of diazinon using Shinwasorb
Column
Silicone OV-11 10% Shinwasorb-U 60/80
Column Size
1.6 m x 3.2 mm I.D. Glass
Column temp.
200℃
Carrier gas
25 mL/min N₂(Adjust so that the retention time of diazinon is about 7 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FID
Injection volume
3.0 μL

View detail


Graph

GC

Samples
Anhydrous caffeine, Fenitrothion in Acetone
Applications
Analysis of Fenitrothion Emulsifiable Concentrate using Shinwasorb
Column
Silicone OV-210 3% Shinwasorb-S 80/100
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
175℃
Carrier gas
51 mL/min N₂(Adjust so that the retention time of Anhydrous caffeine is about 10 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Anhydrous caffeine, Fenitrothion in Acetone
Applications
Analysis of Fenitrothion Microcapsule using Shinwasorb
Column
Silicone OV-210 1.5% Shinwasorb-U 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
165℃
Carrier gas
50 mL/min N₂(Adjust so that the retention time of diazinon is about 17 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Benzyl salicylate, Fenitrothion in Acetone
Applications
Analysis of fenitrothion, phthalthrin mixed emulsifiable concentrate using Shinwasorb -1-
Column
Silicone OV-210 1.5% Shinwasorb-U 60/80
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
175℃
Carrier gas
52 mL/min N₂ (Adjust so that the retention time of fenitrothion is about 15 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Phthalthrin in Acetone
Applications
Analysis of phthalthrin using Shinwasorb ~Isomer ratio phthalthrin~
Column
Silicone DC QF-1 2.5% Shinwasorb-S 80/100
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
200℃
Carrier gas
40 mL/min N₂( Adjusted so that the retention time of 3.Phthalthrin is about 40 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FID
Injection volume
1.0 μL

View detail